2. In Calibration Mode
CH 11
Reason: Resolution(maximum capacity/minimum division) is over the limit
(1/10,000).
Trouble shooting: Lower the resolution in any of the below ways.
Modify maximum capacity in CAL 1 of CAL mode.
Modify minimum division in CAL 2 of CAL mode.
CH 12
Reason: Span setting weight is set under 10% or over 100% of the maximum
capacity.
Trouble shooting: Set span setting weight to 10% ~ 100% range of maximum
capacity in CAL 3 of CAL mode.
CH 13
Reason: Load cell output voltage(span value) is too high at span calibration.
Trouble shooting: Check if weighing platform is in contact with something.
Or set external resolution to lower external resolution.
CH 14
Reason: Load cell output voltage(zero value) is too high or too low at zero
calibration.
Trouble shooting: Confirm that there is nothing on the weighing platform.
Or check if weighing platform is in contact with something.