41
Key Display Description
Key :
Move to the initial
menu
Load the maximum capacity weight and
Press key.
Under SPAN calibration.
Span calibration is terminated.
It is automatically escaped from CAL mode.
Press key.
CAL mode is completed.
Move to the initial menu automatically.
Function: Span Calibration
ZERO=tESt
Tare = Cal
Note 1. “END” message is displayed and move to the initial menu when span
calibration is completed.
Note 2. If the span value is too low or high, “CH 13” will occur.
You lower the resolution and reset the span calibration.
Note 3. After completing CAL mode, it will be moved to initial menu.