Display K e y Platform Description
Step 6
Step 7
Step 8
Step 9
Step
Under zero
calibration.
Zero calibration
is
completed.
It is escaped
from
zero calibration.
CAL mode is
completed.
Move to the
Empty
Empty
Empty
zero=tESt
Tare=Cal
Note 1. If the zero value is too low or high in STEP 6, “CH 13” will occur.
Note 2. If you find “CH 14”, check the platform and the connection of load-cell,
and do this test again.