60
<Checker Mode 5> Relay Operation Graph when #9 of F51 is Set
Step2 (HIGH)
Step1 (LOW)
Judgment Input External Input
SP1 (LOW) Output 1
SP2 (HIGH) Output 2
SP3 (OK) Output 3
Lowest Limit NG Output 4
Upper Limit NG Output 5
Near Zero Output 6
Note.
1. Required set value input: Step 2 > Step 1
2. Near zero output is according to the specified range in F57.
3. T3: Refer to F54 (Delay time of judgment-relay output)
T4: Refer to F55 (Operation time of judgment-relay output)
T5: Refer to F56 (Operation(ON) time of weighing NG relay output)
4.Relay Output
SP1(LOW): During Judgment Input, ON when it is below the value of Step1
SP2(HIGH): During Judgment Input, ON when it is over the set value of Step2
SP3 (OK): During Judgment Input, ON when it is in between Step1 ≤Step2
Lowest Limit NG: ON during Step1 Output, adjust Output Time on T5
Upper Limit NG: ON during Step2 Output, adjust Output Time on T5
Near Zero: F57 Set Value ≥ 0 Range Output
5. Step1`-4(SP 1-4)‘s status lamp in the front panel is operated in the same manner as the RELAY output
T3
T4
T3
T3
T4
T4
T5
T5