45
RS-232C Communication
F23
Function Set the RS-232C Baud Rate
Display Signification
F23. 0 600 bps
F23. 1 1,200 bps
F23. 2 2,400 bps
F23. 3 4,800 bps
F23. 4 9,600 bps
F23. 5 19,200 bps
F23. 6 38,400 bps
F23. 7 57,600 bps
Set-up Range
(0 ~ 8)
F23. 8 115,200 bps
F24
Function Set the RS-232C – Output Format
Display Signification
F24. 0 CAS’s 22 bytes
F24. 1 CAS’s 10 bytes
Set-up Range
(0 ~ 2)
F24. 2 18-byte format (AND, FINE)
F25
Function Set the RS-232C - Output Mode
Display Signification
F25. 0
Use BCC (ERROR CHECK) in the Extended Command
Mode
F25. 1 Transmit when stable / unstable (Stream Mode)
F25. 2 Transmit only when weight is stable
F25. 3 Simple Command Mode
F25. 4
Extended Command Mode
Do not use BCC (ERROR CHECK)
F25. 5
Transmit weight value when ID number is entered (KEY)
Change to ID number value at the location of (CAS 22 BYTE)
6,7 BYTE and transmit
F25. 6
Use the Indicator as the External Indicator. Use 18BYTE
Communication Format
( C I-502A/505A GND2,RXD2 Connection)
Set-up Range
(0 ~ 7)
F25. 7
Transmit only when weight is stable(F32.1 set to Automatic
Print)
Transmit only when the Print key is pressed
Note 1. When setting F25 to 3, refer to the <Simple Command Mode Table> below.
Note. When setting F25 to 0 or 4, refer to the appendix 2 <How-to-Use the Extended Command Mode>.