8
Technical Specification
Analog Part & A/D Conversion
Input sensitivity
0.45 / D
Adjustment range of Zero - 0.6mV + 42.0mV
Load Cell Excitation Voltage DC 10V (±5V)
Temperature Coefficient
Zero: ±20 PPM /
SPAN : ±20 PPM /
Input Noise
±0.6 P. P
Input Impedance
Over 10
A/D conversion method ∆∑
A/D resolution 520,000 Count(19bit)
A/D conversion rate 200 times / sec
Nonlinearity 0.01% FS
Digital Part
Span Calibration
Full Digital Calibration
(Set one automatic weight calibration)
Display
7Segment 3 red-colored digits FND 2EA
CI-502A Character size : 20.0(H) X 11.0(W) mm
CI-505A Character size: 14.1(H) X 8.1(W) mm
Key Function key (4EA), Number and Function key (12 EA)
Division X1, X2, X5, X10, X20, X50
Below Zero Indication Negative “-“ sign
Allowed Tare Range Full capacity
Status lamp(lamp) STABLE, ZERO, TARE, NET, SP1, SP2, SP3, SP4, COMM
General Specifications
Power SMPS Free Voltage AC 100V ~ 250V , 50/60 Hz
Product Size 195(W) x 189(D) x 98(H)
Operating Temperature -10 ~ +40℃℃
Product Weight Approx.1.8 kg
Fuse Capacity 6A L250V
Power Consumption Approx. 10W