16. CHECK MESSAGE
(1) Operating Mode
CH 01
Reason
Internal RAM is erased.
Solution
Confirm the setting value in SET Menu.
CH 02
Reason
Load cell connection failure or error in A/D conversion part.
Solution
Check the Load Cell connector to see if the polarity of signal is reversed.
CH 03
Reason
stable to initialize the scale.
Solution
Place the scale on a stable surface.
CH 04
Reason
Zero range deviate from the set range (±10% of Max.)
Solution
Confirm that there is nothing on the weighting platform.
Over
Reason
The display weight is larger than the Maximum Capacity you've set.
Solution
Don't load the article whose weight is larger than the Max. Capacity on the
platform scale. This may damage Load Cell.