4. The Features of EXP-2000A
(1) Features
High quality, High accuracy
Appropriate for weight and measurement system
Easy operation
RFI/EMI screened
WATCHDOG circuitry (System restoration)
WEIGHT BACK-UP (Weight memory at sudden power failure)
(2) Main Function
Store date, time and calculated data at sudden power failure
Adjustable display rate(Digital filter function)
Printer connection (serial printer)
Tare weight setting with keys
Storage of measured times
PC interface
Zero-tracking function
Automatic, manual hold function
Users can set the desirous max. weight and a division freely
Print date and time by built-in clock
Self hardware test
5. Technical Specification
Analog Part & A/D Conversion
Loadcell Exciting Voltage DC 5V, Up to 4×350 load cells
Full Scale Input Signal 20 mV, including dead load
Zero Adjust Range
0.05 mV 5 mV
2 /D (H-44, OIML)
Input Sensitivity
0.5 /D (Non H-44, OIML)
System Linearity
0.02 of FS
A/D internal resolution 1 / 200,000
5,000 dd (H-44, OIML)
A/D external resolution
30,000 dd (H-44, OIML)
A/D conversion speed 10 times/sec