12. CALIBRATION MODE
(1) How to enter
Turn on the power while pressing the key on the front of the indicator.
(2) CALIBRATION MENU ( CAL 1 - CAL 5 )
CAL 1 : Maximum Capacity
CAL 2 : Minimum Division
CAL 3 : Setting Weight
CAL 4 : Zero Calibration
CAL 5 : Span Calibration
CAL 1 (press 'SET' key to move to CAL 1)
FUNCTION : Maximum Capacity SET
RANGE ---> 1 ~ 99,999
KEY LED DISPLAY DESCRIPTION
SET key : store and move into
next menu
NUMERIC key : set value change
CLR key : initial value
5000
20000
5000 kg
20000 Kg
REF 1. The maximum capa. means the maximum weight that scale can measure.
REF 2. Do not input the resolution, there is no need to input the resolution
which is automatically calculated.
CAL 2
FUNCTION : Minimum Division SET
RANGE ---> 0.001 ~ 500
KEY LED DISPLAY DESCRIPTION
SET key : store and move into
next menu
NUMERIC key : set value change
CLR key : input the point, initial
value.
1
0.2
0.05
0.001
1 kg
0.2 kg
0.05 kg
0.001 kg
REF 1. The minimum division means the value of one division.
REF 2. External resolution is obtained by dividing the min. division by the
maximum capacity. Set the resolution to be within 1/10,000.
REF 3. When you press other keys except '1', '2', '5' and '0', A retry beep will sound.
8