APPENDIX
EN
|
Intelect
®
Mobile 2 ELECTROTHERAPY User Manual
ELECTROMAGNETIC COMPATIBILITY EMC TABLES CONTINUED
Recommended separation distances between portable and mobile RF communications equipment and the Intelect
®
Mobile
The Intelect
®
Mobile is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the Intelect
®
Mobile can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and
the Intelect
®
Mobile as recommended below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
P (W)
Separation distance according to frequency of transmitter
d (m)
kHz to MHz
d =
[
3,5
]
√P
V
1
(where V = V)
MHz to MHz
d =
[
3,5
]
√P
E
1
(where E = V/m)
MHz to , GHz
d =
[
7
]
√P
E
1
(where E = V/m)
, , , ,
, , , ,
, , ,
, , ,
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reflection from structures, objects and people.