APPENDIX
EN
|
Intelect
®
Mobile 2 ELECTROTHERAPY User Manual
ELECTROMAGNETIC COMPATIBILITY EMC TABLES CONTINUED
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The Intelect
®
Mobile is intended for use in the electromagnetic environment specified below.
The customer or the user of the Intelect
®
Mobile should assure that it is used in such an environment.
Immunity Test IEC Test Level Compliance Level Electromagnetic Environment - Guidance
Conducted RF IEC
--
Radiated RF IEC
--
Vrms kHz to MHz
V/m MHz to , GHz
[V] V, where V = V
[E] V/m, where E = V/m
Portable and mobile RF communications
equipment should be used no closer to any part of the Intelect
®
Mobile ,
including cables, than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
d =
[
3,5
]
√P
V
1
d =
[
3,5
]
√P 80 MHz to 800 MHz
E
1
d =
[
7
]
√P 800 MHz to 2,5 GHz
E
1
where P is the maximum output power rating of the transmit- ter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic
site survey,a should be less than the compliance level in each frequency range.b
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot
be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the
Intelect
®
Mobile is used exceeds the applicable RF compliance level above, the Intelect
®
Mobile should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the Intelect
®
Mobile .
b
Over the frequency range kHz to MHz, field strengths should be less than [V] V/m.