APPENDIX
EN
|
Intelect
®
Mobile 2 ELECTROTHERAPY User Manual
ELECTROMAGNETIC COMPATIBILITY EMC TABLES CONTINUED
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The Intelect
®
Mobile  is intended for use in the electromagnetic environment specified below.
The customer or the user of the Intelect
®
Mobile  should assure that it is used in such an environment.
Immunity Test IEC  Test Level Compliance Level Electromagnetic Environment - Guidance
Conducted RF IEC
--
Radiated RF IEC
--
 Vrms  kHz to  MHz
 V/m  MHz to , GHz
[V] V, where V = V
[E] V/m, where E = V/m
Portable and mobile RF communications
equipment should be used no closer to any part of the Intelect
®
Mobile ,
including cables, than the recommended separation distance calculated from the
equation applicable to the frequency of the transmitter.
Recommended separation distance
d =
[
3,5
]
P
V
1
d =
[
3,5
]
P 80 MHz to 800 MHz
E
1
d =
[
7
]
P 800 MHz to 2,5 GHz
E
1
where P is the maximum output power rating of the transmit- ter in watts (W)
according to the transmitter manufacturer and d is the recommended separation
distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic
site survey,a should be less than the compliance level in each frequency range.b
Interference may occur in the vicinity of equipment marked with the following
symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is aected by absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot
be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which the
Intelect
®
Mobile  is used exceeds the applicable RF compliance level above, the Intelect
®
Mobile  should be observed to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the Intelect
®
Mobile .
b
Over the frequency range  kHz to  MHz, field strengths should be less than [V] V/m.

Table of Contents

Other manuals for Chattanooga Intelect Mobile 2

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chattanooga Intelect Mobile 2 and is the answer not in the manual?

Chattanooga Intelect Mobile 2 Specifications

General IconGeneral
BrandChattanooga
ModelIntelect Mobile 2
CategoryMedical Equipment
LanguageEnglish

Related product manuals