Description of Panel
4-9
BINNING:
Binning setting. Press [F2] to switch OFF/ON. If setting is “ON” it will show “BIN X” word
under measurement screen. This parameter is set under MAIN INDEX menu.
TEMP.:
It includes the functions of temperature measurement and correction. There are three
types of selections OFF, AUTO and +20°C. Press [F3] for switch under the third page,
+20°C is the value of inputted manually. The actual display value depends on the main unit,
+20°C is an example so the shown numeral isn’t always 20. The detailed operations
please see 8.3.2 Operating Description.
UNDCNA:
It should coordinate with the setting for use (the necessary setting) under MAX
measurement speed and DC+ NA test mode. Press [F1] to switch Off/ON under the
screen.
<<There is a symbol “
” on upper left side, when measuring it will turn clockwise
repeatedly, and its turning speed is changed by measuring speed FAST/ MEDIUM/
SLOW. Trigger once the symbol will turn once under manual trigger. External
trigger is the same as manual trigger. >>
4.4.2 COMPARE Setting
1. Power on the instrument after all are normal, enter main index by pressing [MAIN INDEX]
as below shown.
← High, low limit compare test
← Temp. conversion setting
Press [F1] to enter compare setting screen as below shown:
Enter “COMPARE SET” as below shown.
P A R E S E T >
D I G I T U P
N O
I N A L : 0 0 0
. 0 0 0 0
Ω
U P
P E R : 0 0 0
. 0 0 0 0
Ω D I G I T D O
E R : 0 0 0
. 0 0 0 0
Ω
D I G I T
P r e s s
A I N I N D E X t o e x i t
← The cursor positioned number digit
up.
← The cursor positioned number digit
down.
← Set high/low limit value is ABS
(absolute value) or % (percentage).
Setting example:
Assumed NOMINAL is set to 100mΩ, please follows the below method