Milliohm Meter 16502 User’s Manual
4-10
to operate:
(1) Press [F4], set MODE to percentage (%).
(2) Press [F3] to move the cursor to NOMINAL position, and move the cursor to hundred
then press [F1] key to adjust the number to 1. Press [F3] key to move the cursor to
(-) position, then press [F1] key to set unit to m.
(3) Press [] to move the cursor to UPPER position, at this time the cursor stay at
hundred then press [F3] to move the cursor to the first decimal place then press [F1]
(digit up) to adjust the number to 1.
(4) After the last setting is completed, press [TRIGGER] key then the cursor will move to
the next item – LOWER, the setting is -001.0000%.
(5) After the setting is completed the menu should be as the following:
P A R E S E T >
D I G I T U P
N O
A I N I N D E X t o e x i t
After the setting is completed, press [MAIN INDEX] to exit. Press [MEAS DISPLAY] key
back to page 3 of measurement menu, select [F1] key to enable COMPARE function for
setting ON, ON-⊿% and ON-⊿.
As the following menu:
E A S . D I S P L A Y >
*
C O
⊿
B I N N I N G : O F F
T E
P . : O F F
N E X T P A G E 3
← Select ON, ON-⊿% and ON-⊿
Please set warning sound of COMPARE function under SOUND MODE of SYSTEM
CONFIG screen.
4.4.3 BINNING Setting
1. Power on the instrument after all are normal, enter main index by pressing [MAIN INDEX]
as below shown.
Press [F2] to enter binning setting screen as below shown: