Return: The query returns CLEAR or HOLD.
(6) Command: SYSTem:CONTrast < NR1>
Function: It sets or queries the contrast of LCD.
Parameter: <NR1> 0 ~ 15
Return: The query returns a numeric value in the format NR1.
(7) Command: SYSTem:KLOCk {ON(1)|OFF(0)}
Function: It sets or queries if the key of 16502 is locked.
Parameter: {ON(1)|OFF(0)}
Return: The query returns <NR1> 0 or 1.
(8) Command: SYSTem:PRESet
Function: It resets the 16502 to its default state.
Parameter: None
Return: None
The default state is as below table.
Item Default Value :SYST:PRES *RST
Saved
Position
DRIVE
PULSE+/- PULSE+/-
Saved
value
EEPROM
DRY
OFF OFF
Saved
value
EEPROM
TRIG INT
Not be
influenced
Not be
influenced
EEPROM
RANGE MODE
AUTO AUTO
Saved
value
EEPROM
RANGE
2M 2M
Saved
value
EEPROM
SPEED
FAST FAST
Saved
value
EEPROM
ZERO
OFF OFF
Saved
value
EEPROM
COMPARE
OFF OFF
Saved
value
EEPROM
BINNING
OFF OFF
Saved
value
EEPROM
TEMP.
OFF OFF
Saved
value
EEPROM
NUDCNA
OFF
OFF
OFF
-
The nominal setting of
comparator
0.000 0.000
Saved
value
EEPROM
The high limit setting of
comparator
0.000 0.000
Saved
value
EEPROM
The low limit setting of
comparator
0.000 0.000
Saved
value
EEPROM
The showing mode setting
of comaprator
ABS ABS
Saved
value
EEPROM
The bin sort parameter
0.0000 0.0000
Saved
value
EEPROM
Temp digit unit
°C °C
Saved
value
EEPROM
The initial resistance of
temp conversion
Saved
value
EEPROM