EasyManua.ls Logo

Chroma 66203 - Limit

Chroma 66203
146 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation
4-19
4.6.1 Limit
Figure 4-28 Limit Function Setup Screen
Enable the Limit function by setting the detection time and the upper/lower limits of each
parameter can judge if the measured data is within the specification. The parameters
include V, Vpk+, Vpk-, I, Ipk+, Ipk-, Is, W, PF, VA, VAR, CFi, VTHD, ATHD, E and F. The
user can select one or several of them for setting.
Following is the procedure to set the parameter:
1. Press SETUP and select Limit. The Limit indicator is on indicating it is in Limit menu.
2. Use , to select on in the 1
st
display window and press TRIG/ENTER to confirm it.
3. Use , to set the time in the 2
nd
display window and press TRIG/ENTER to enter into
the parameter setting screen.
4. Use , to select the desired testing channel in the 1
st
display window. If the
parameter upper and lower limits of every channel are the same, select all and press
TRIG/ENTER to confirm it.
5. Use , to select the test parameter in the 2
nd
display window and press Enter/Trig to
confirm it.
6. Use , to set the upper limit in the 3
rd
display window. If there is no upper limit, set
“-----” and press TRIG/ENTER to confirm it.
7. Use , to set the lower limit in the 4
th
display window. If there is no lower limit, set
“-----” and press TRIG/ENTER to confirm it.
8. Repeat step 4~7 to set other test channel, parameters and upper/lower limits. When
done, press SETUP to exit the menu and the GO/NG indicator will turn on.
9. Press TRIG/ENTER to begin the GO/NG testing on every channel. The GO/NG will
blink during the test.

Table of Contents

Related product manuals