Using Remote Control
5-11
5.7.2 Instrument Commands
SYSTEM Sub-system
SYSTem:ERRor?
Description: This command queries the error string of the command parser.
Setting syntax: none
Setting parameters: none
Setting example: none
Query syntax: SYSTem:ERRor?<PMT>
Return parameters: <SRD>,
0,“No Error”
1,“Data Format Error”
2,“Data Range Error”
3,“Command Error”
4,“Execution Error”
5,“Too many Errors”
Header on: :SYSTEM:ERROR<space><SRD><RMT>
Header off: <SRD><RMT>
Query example: none
SYSTem:HEADer
Description: This command turns response headers ON or OFF. The defult is
OFF.
Setting syntax: SYSTem:HEADer<space><CRD><PMT>
Setting parameters: <CRD>, ON | OFF
Setting example: none
Query syntax: SYSTem:HEADer?<PMT>
Return parameters: <CRD>, ON | OFF
Header on: :SYSTEM:HEADER<space><CRD><RMT>
Header off: <CRD><RMT>
Query example: none
SYSTem:TRANsmit:SEParator
Description: This command sets the message unit separator for response
messages. The default is 0(Comma).
Setting syntax: SYSTem:TRANsmit:SEParator<space><NR1><PMT>
Setting parameters: <NR1>, 0 ~ 1; 0 : Comma(,) 1 : Semicolon(;)
Setting example: none
Query syntax: SYSTem:TRANsmit:SEParator?<PMT>
Return parameters: <NR1>, 0 ~ 1
Header on: :SYSTEM:TRANSMIT:SEPARATOR<space><NR1><RMT>
Header off: <NR1><RMT>
Query example: none
SYSTem:TRANsmit:TERMinator
Description: This command sets the data terminator for response messages.
The default is 0(LF).
Setting syntax: SYSTem:TRANsmit:TERMinator<space><NR1><PMT>
Setting parameters: <NR1>, 0 ~ 1; 0 : LF 1 : CR+LF

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Chroma 66203 and is the answer not in the manual?

Chroma 66203 Specifications

General IconGeneral
Model66203
CategoryMeasuring Instruments
Channels3
Harmonic MeasurementUp to 50th order
DisplayLCD
Current Accuracy±(0.1% reading + 0.1% range)
InterfaceGPIB

Related product manuals