Digital Power Meter 66203/66204 User’s Manual
5-12
Setting example: none
Query syntax: SYSTem:TRANsmit:TERMinator?<PMT>
Return parameters: <NR1>, 0 ~ 1
Header on: :SYSTEM:TRANSMIT:TERMINATOR<space><NR1><RMT>
Header off: <NR1><RMT>
Query example: none
SYSTem:VERsion?
Description: This query returns an <NR2> formatted numeric value
corresponding to the SCPI version number for which the instrument
complies.
Setting syntax: none
Setting parameters: none
Setting example: none
Query syntax: SYSTem:VERsion?<PMT>
Return parameters: <SRD>, 1991.1
Header on: :SYSTEM:VERSION<space><SRD><RMT>
Header off: <SRD><RMT>
Query example: none
SYSTem:LOCal
Description: This command can only be used under control of USB. If
SYST:LOC is programmed, the Power Meter will be set in the
LOCAL state, and the front panel will work.
Setting syntax: SYSTem:LOCal<PMT>
Setting parameters: none
Setting example: none
Query syntax: none
Return parameters: none
Query example: none
SYSTem:REMote
Description: This command can only be used under control of USB. If
SYST:REM is programmed, the Power Meter will be set in the
REMOTE state, and the front panel will be disabled except the
<SETUP>key pressed.
Setting syntax: SYSTem:REMote<PMT>
Setting example: none
Setting parameters: none
Query syntax: none
Return parameters: none
Query example: none
STATUS Sub-system
STATus:QUEStionable[:EVENt]?
Description: This query returns the value of the Questionable Event register. The
Event register is a read-only register which holds all events that are
passed by the Questionable NTR and/or PTR filter. If QUES bit of
the Service Request Enable register is set, and the Questionable