18
AFQm
Instruction Manual
300
500
65
65
15
15
Fixing screws
Fixing screws
Figure 5: Installation Wall-type AFQm.
Use 4 M8 fixing screws.
The ventilation grilles must not obstructed or covered at any time.
3.4.2.- RACKTYPE AFQm
The Rack-type AFQm should be installed in a 19” rack cabinet.
The height of the device is equivalent to 9U. (U is the rack unit, 1U = 4.445 cm)
You can install more than one AFQm in a single cabinet.
Follow these steps to install it inside the cabinet:
1.- Extend the rack cabinet’s anti-tip feet.
Unless the rack cabinet is secured to the floor, the anti-tip feet have to be extended and
fixed to the floor, to ensure maximum safety during the assembly process.
2.- Open or remove the door at the front of the cabinet.
3.- Place the AFQm on the cabinet’s rails or shelves. Make sure that they are suitable for the weight of
the device; use cross braces if necessary.
Do not use front handles to carry the module.