- 15 -
10.4. Defition temperature of drying
The temperature can be set in the range from 50ºC to 160ºC.
Function keys Applications
Change of value of current flashing
nubmer.
Values are changed by pressing key
many times in range of 0-9
Choose number which value must be
defined. Every pressing key mores cursor
in right side.
confirm value. After pressing key F next
parameter is shown automatically.
Attention: If settled temperature of drying is lower then temperature
of surrounding moisture balance will not light halogen lamps on.
10.5. Definition time of testing samples
The STROBE INTERVAL time is the time between one set of
results and the next set of results. The interval time sets how often
the current results are output to the RS-232 interface. When using
MODE 1, 2 or 3 the test will stop when 3 results are the same,
indicating that there is no more moisture to be taken from the
sample.