74 P2077SB/EN 2011-09 77l_Fehlersuche-en.fm, 06.09.2011
Troubleshooting
15
Measurement card:
Sampling clock from servo:
NOK
The system clock from the servo
amplifier is missing
Internal error
➔ Replace the TSE/TUSE
Measurement card:
Servo type correct: NOK
The selected servo amplifier type
is incorrect
Error in programming
➔ Check the system program-
ming
Self-identification of transducer is
not OK
Check the transducer
➔ Replace the transducer
Internal error
➔ Replace the TSE/TUSE
Measurement card:
Servo par. matching servo:
NOK
The parameter set selected from
the measuring board is not
present in the TSE/TUSE.
Error in programming
➔ Check the system program-
ming
Self-identification of transducer is
not OK
Check the transducer
➔ Replace the transducer
Internal error
➔ Replace the TSE/TUSE
Measurement card:
ARCNET communication:
Dup-ID
The same ARCNET addresses
are preset
Several TSE/TUSE are preset to
the same ARCNET address
Check the preset ARCNET
addresses
➔ Set different addresses
Measurement card:
ARCNET communication:
Recon
The ARCNET is temporarily
disrupted
ARCNET terminator missing
➔ Plug in the ARCNET termina-
tor
No power supply to the ARCNET
terminator
➔ Switch on the power supply
for the last device
Error in cabling
➔ Plug in all cables and lock
Error in potential equalization
➔ Connect potential equaliza-
tion cable
Internal error
➔ Replace the TSE/TUSE
Measurement card:
Initialization: NOK
Initialization error in measuring
board
Internal error
➔ Replace the TSE/TUSE
➔ Inform DGD service
Measurement card:
Flash image: NOK
Flash image in the measuring
board is not OK
The incorrect program has been
transmitted by mPro400GC-M
➔ Check the program version
The program transfer was
interrupted
➔ Repeat the program transfer
Internal error
➔ Replace the TSE/TUSE
Measurement card:
Voltage +3.3 V (…): NOK
The +3.3 V supply to the measur-
ing board is outside the limits of
+3.24 V... +3.53 V
The power adapter for the inter-
nal supply to the measuring
board is overloaded or defective.
Internal error
➔ Replace the TSE/TUSE
Error
Description
Possible causes Measures and
remedies