74 P2077SB/EN 2011-09 77l_Fehlersuche-en.fm, 06.09.2011
Troubleshooting
15
Measurement card:
Sampling clock from servo:
NOK
The system clock from the servo
amplifier is missing
Internal error
Replace the TSE/TUSE
Measurement card:
Servo type correct: NOK
The selected servo amplifier type
is incorrect
Error in programming
Check the system program-
ming
Self-identification of transducer is
not OK
Check the transducer
Replace the transducer
Internal error
Replace the TSE/TUSE
Measurement card:
Servo par. matching servo:
NOK
The parameter set selected from
the measuring board is not
present in the TSE/TUSE.
Error in programming
Check the system program-
ming
Self-identification of transducer is
not OK
Check the transducer
Replace the transducer
Internal error
Replace the TSE/TUSE
Measurement card:
ARCNET communication:
Dup-ID
The same ARCNET addresses
are preset
Several TSE/TUSE are preset to
the same ARCNET address
Check the preset ARCNET
addresses
Set different addresses
Measurement card:
ARCNET communication:
Recon
The ARCNET is temporarily
disrupted
ARCNET terminator missing
Plug in the ARCNET termina-
tor
No power supply to the ARCNET
terminator
Switch on the power supply
for the last device
Error in cabling
Plug in all cables and lock
Error in potential equalization
Connect potential equaliza-
tion cable
Internal error
Replace the TSE/TUSE
Measurement card:
Initialization: NOK
Initialization error in measuring
board
Internal error
Replace the TSE/TUSE
Inform DGD service
Measurement card:
Flash image: NOK
Flash image in the measuring
board is not OK
The incorrect program has been
transmitted by mPro400GC-M
Check the program version
The program transfer was
interrupted
Repeat the program transfer
Internal error
Replace the TSE/TUSE
Measurement card:
Voltage +3.3 V (…): NOK
The +3.3 V supply to the measur-
ing board is outside the limits of
+3.24 V... +3.53 V
The power adapter for the inter-
nal supply to the measuring
board is overloaded or defective.
Internal error
Replace the TSE/TUSE
Error
Description
Possible causes Measures and
remedies