CB-MF-SOP-002
A/1
28
1. recognition accuracy
The analyzer automatically identify, and the negative and positive coincidence rate between the
single result and microscopy were: red blood cells ≥ 70%, white blood cells ≥ 80%, and cylinder
≥ 50%.
The false negative rate of the analyzer is no more than 3%.
2. Repeatability
The coefficient of variable CV of counting result for the analyzer is when the cell concentration is
50 cells/μl, CV≤25%; when the cell concentration is 200 cells/μl, CV≤15%.
3. Stability
Within 8 hours of starting the analyzer, the coefficient of variation of cell count is CV≤15%.
4. Contamination rate
The contamination rate of cells by the analyzer should be ≤0.05%.
5. Background Count
Analyzer blank counts for WBC and RBC should be 0.
6. Detection Limit
The analyzer can detect the sample of red blood cells and white blood cells with the
concentration level of 5 / μl.
9 Support
9.1 Maintenance Statement
Only trained professionals are qualified to maintain the instrument.
Only the original parts can be used for repair and replacement.
Before performing any instrument maintenance operation, including instrument transportation,
the power should be turned off and the power cord of the main engine should be unplugged.
The manufacturer reserves the right to make final changes, so the description of this manual
may be slightly different from the actual situation.
Updated documentation of specific versions of the instrument is available from the
manufacturer.
9.2 Order Information
Consumables:
COBIO series counting chamber
Accessories of S80:
D1-3+D003-4 Power cord for main engine
1394A 线 1.5 米 Image acquisition firewire cable
Gigabit Ethernet cable 1.5m
RS232(公母) 1.5 米 Connecting cable
T6.3AL250V Fuse (power socket)
T3.15AL250V Fuse (control circuit board, power box)
Liquid level alarm switch assembly