CB-MF-SOP-002
A/1
30
Storage Temperature: -20-40℃
Storage Humidity: <93%
Atmospheric Pressure: 760hPa~1060hPa
Bar-code Scanner:
Barcode identification type:
One-dimensional: code39\Codabar\code128\UPC\EAN, etc.
Two-dimensional: QR Code\Data Martix, etc.
Scanning peak wavelength of light: 451nm, 575nm
Maximum laser power: <0.39mW
Safety of laser products - part 1: equipment classification and requirements, this instrument
belongs to Class Ⅰ laser product.
Test Tube rack
Use the test tube rack of original factory only.
Test Tube
Minimum sample size of the test tube: 2ml
Urine homogenization treatment: Absorption and spitting mixed
Height(conical bottom tube): 100 – 110 mm
Height(Flat bottom tube): 100 – 105 mm
Tip diameter of test tube: 16 - 17.5 mm
Maximum diameter on top of test tube: 16.5 mm
Counting Chamber (consumables)
Package: 50pcs/container
Maximum counting chamber load: 600pcs (12 containers)
Sample size: 175μL
Centrifuge:
Centrifugal speed: 2000RPM, Centrifugal time: 10s
The centrifugal liquid can only be urine samples, water and related quality control products, etc.
If the user lets the equipment absorb other unknown samples, the consequences will be their own.
The main material of the centrifugal arm module was POM, the density is about 1.41g /cm
3
, the
embedded metal material is S136, the density is about 7.85g /cm
3
.
The overall mass of the centrifugal arm assembly is about 35g.
The overall volume of the centrifugal arm module is about 13 cm
3
.
Fixing method of centrifugal arm assembly: Locate by the locating pin and mounting seat of
centrifugal arm, and fixing by magnetic force of 2 magnetic steel on mounting seat of centrifugal arm.
Light Source:
Green