4.1 Calibration for Toxic sensors and Oxygen sensors
4.1.1 Calibration Zero level (ZERO calibration)
Proceed as follows:
Activate the USER MENU as described in chapter 3.2.2.
Select CAL ZERO menu and follow the procedure shown below:
TEXT: [IN PROGRESS]
BLINKS
IMPORTANT NOTE:
TOXIC SENSORS
o If the gas detector is mounted in
an area with clean air the ZERO
calibration can be started.
o If the surrounding environment
does not contain clean air,
synthetic air should be used
during the ZERO calibration
o Apply synthetic air with a flow of
20-30l/h
IMPORTANT NOTE:
OXYGEN SENSORS
o Always apply Nitrogen (N
2
) to
perform a ZERO calibration, flow
20-30l/h.