EasyManuals Logo

Consilium ST650EX User Manual

Consilium ST650EX
49 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #3 background imageLoading...
Page #3 background image
ST650EX / ST650-CO
2
/ ST350EX
v3.16
3
IMPORTANT NOTES:
- Read and understand this user manual and all other safety instructions before using this equipment.
- This equipment must be operated by trained and experienced personal only.
- If the ST650EX / ST650-CO
2
/ ST350EX is used in a hazardous environment, the relevant regulations for
explosion hazardous equipment must be observed at all times.
- The gas detector should be installed in an area that is free from shocks or vibrations.
- The gas detector should be accessible for maintenance and/or calibration purposes at all time.
- The gas detector (sinter) should be pointed downwards to avoid accumulation of moisture, dirt, dust and/or
chemical residue. It is also allowed for indoor applications to point the sensor horizontal as long as the
sensor is not affected by moisture, dirt, dust and/or chemical residue.
- IP ratings do not imply that the equipment will detect gas during and after exposure to those conditions.
- During installation the gas detector should be isolated from local earth or conductive surface. The gas
detector should only be connected to a clean earth through cable gland and the shielding of the signal
cable.
- Avoid signal cables close to high power cables, inverters, radio transmission lines, computer
communication cables or cables carrying pulses of high current.
- The ST650EX is designed for the detection (temporarily) of unexpected gas concentrations. The
ST650EX measures various Hydrocarbons. Other gases than the target gas (for example 0-100% LEL
Methane or 0-100% LEL Propane) gives a deviant reading.
- The ST650EX equipped with a catalytic bead sensor is not suitable for use in environments where
continuous concentrations of gas are present. Continues concentrations of gas affects the life time of the
sensor and creates a significant loss of sensitivity.
- A gas detector provided with a catalytic bead sensor is not suitable in areas where silicon compounds,
organometallic compounds, organic phosphate esters, halogen-containing hydrocarbons and sulphur
compounds are present. These environmental factors may significantly reduce the life of the sensor.
- Gas detectors with a catalytic bead gas sensor requires > 10% Oxygen for a correct sensor reading.
- To minimize deviations, circumstances during calibration should approach normal operational
environment (temperature, humidity and pressure).
- Never change a sensor in a potentially hazardous atmosphere while the transmitter is still powered. Keep
cover tight when circuits alive.
- When a combustible gas sensor exceeds the maximum range, the gas detector should be re-calibrated.
- Readings between -3 and +3 are shown as [0] (4mA), readings < -3 are shown as [-0] (2,4…3,52mA),
readings <-10% f.s. are shown as SF (output fixed 2,4mA), out of range is shown as [100] (blinking) and
output is fixed 22mA (will stay at 22mA until unit is recalibrated).
- The time of response of the entire system is determined by the time of response of all parts of
equipment within the gas detection system.
- The EU type examination certificate (EN 60079-29-1) only applies to Methane and Propane for Catalytic
Bead sensors and NDIR sensors in range of 100% LEL.
- IEC60079-20-1 can be used for converting test and calibration gas concentrations from % LEL to
% volume fraction.
- For calibration and maintenance; we recommend to read the procedures described in IEC 60079-29-2
for reference.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Consilium ST650EX and is the answer not in the manual?

Consilium ST650EX Specifications

General IconGeneral
BrandConsilium
ModelST650EX
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals