EasyManua.ls Logo

Cosman G4 - Page 38

Cosman G4
115 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cosman G4 Radiofrequency Generator Service Manual (SPI 11291 Rev A)
9.1.1 Additional Example Screens
The One Touch Stimulation screen will appear differently depending on the current settings:
Electrode Setup settings (see Section 7.1)
Stimulation Mode: Sensory or Motor
Auto Ramp Setting: Off/Manual or On/Auto (see Section 9.1.2)
Control Units setting: Voltage- or Current-controlled output (see Section 9.1.2)
Sensory
Stimulation Mode
Two Bipolar Pairs
Sensory
Stimulation Mode
One Electrode,
A
uto Ramp = On &
U
ni
ts
=
Cu
rr
e
n
t
Bipolar readings:
Voltage/Current and Impedance
between the two electrodes in
the bi
p
olar
p
ai
r
Motor Mode
Indicator
Motor Stimulation
Mode
Three Electrodes
Mode: Select Sensory
or Motor Stimulation
output
Settings Panels:
Mode switches
between Sensory &
Motor settings
Electrode Panel (Black area):
Displays all readings for the
indicated bipolar electrode pair,
e.g. “Electrode 1+2”.
Temperature Readings:
One Temperature reading
from each electrode in
bipolar pair. For instance,
“E1” indicates a reading
from Electrode 1.
Electrode Select Buttons:
Select between bipolar
electrode pairs 1+2 or 3+4.
A
uto Ramp & Units
Indicator: Reads “Auto” if
A
uto Ramp=On, and
“Manual” otherwise.
Reads “Voltage” or
“Current” to match the
Control Units setting.
Output Level Reading:
Shown here in units of
Current (mA). This is
determined by the
Control Units = Current.
Output Increment &
Decrement Buttons:
A
bsent if Auto Ramp = On.
Electrode Select Buttons:
A
bsent when Electrode Setup
for one electrode or one
bipolar pair.
Section 9

Table of Contents