the gauge will prompt before every reading to manually enter
a depth.
Under this function, you can calibrate the depth sensors. In a
rare case, they may fall out of calibration.
Under Auto Depth function, select 1 to Enable/Disable the Auto
Depth, or 2 for Calibration and follow the screen prompts.
9. Review Standard (STD) Counts
The Elite gauge stores the last 30 standard counts and the date
of measurement. Use this function and the UP or DOWN key to
scroll through all the standard counts. A record of this data can
be downloaded by inserting a USB drive in the gauge and
pressing the STORE key. The standard counts will be stored in
the USB drive for future use.
10. Select Language
Elite is currently designed to work in two languages, English and
Spanish. (Press 1 then 0) Select 1 for English and 2 for Spanish.
11. Set Units
This menu function allows the gauge to set the default units to
pounds per cubic feet (PCF or lb./ft3), kilograms per cubic
meter (kg/m3), or grams per cubic centimeter (g/cc). The
MA/PR will need to be entered in the same units as the default
selection. Depth will be based on the units you select: either
inches for lb./ft3 or millimeters for kg/m3 or g/cc.
To set this option:
1. Press MENU.
2. Press the 11 button (1 then 1).
3. Select PCF, kg/m3, or g/cc with number keypad.