XL™ Series SDI-12 Operations 14-21
Response to a "aM!" command: (when the system is set to factory defaults)
Command Response Time Values Description
------------- --------------------- ------- -------- -------------------------------
"aM!" "a0063<cr><lf>" 6 sec 3 Initiate pressure and temperature
measurement
Subsequent Command Response
--------------------------- ------------------------------------
"aD0!" a+xxx.xxx+yyy.yyy+zz.z<cr><lf>
Key Description Units
------------- ------------------------- -----------------------------------
xxx.xxx Current Stage Feet (Default)
yyy.yyy Current Pressure Pounds-per-Square Inch (PSI)
zz.z Current Temperature Centigrade (EC)
Note: The number of parameters returned and their order, can be modified by the user.
Response to a "aM1!" command:
Command Response Time Values Description
------------- --------------------- ------- -------- -----------------------------
"aM1!" "a0084<cr><lf>" 8 sec 4 Initiate diagnostic pressure and
temperature measurement
Subsequent Command Response
--------------------------- --------------------------------------------------------
"aD0!" a+xxx.xxx+yy.y+XXXXXX+YYYYYY<cr><lf>
Key Description Units
------------- ------------------------- -----------------------------------
xxx.xxx Current Pressure Pounds-per-Square Inch (PSI)
yy.y Current Temperature Centigrade (EC)
XXXXXX Current Pressure Raw Reading
YYYYYY Current Temperature Raw Reading
H-350XL™ Supported Measurement Commands:
14.7.6 Initiate Verify Command
The Verify Command causes a verify sequence to be performed. The result of this command is