XL™ Series Scanning Options 9-3
PC Interface View Built In Display Screen
R - Scan Rate: HH:MM:SS
Scan Rate [HH:MM:SS]
This screen displays the current scan rate. The scan rate can be set in the range of 1 second to 24
hours. If the scan rate is set faster than the time the XL™ takes to make one scan then the actual
scan rate will be a multiple of the entered value. For example a 30 second scan rate will
automatically turn into a 1 minute scan rate if the measurement process takes 35 seconds to
complete.
PC Interface View Built In Display Screen
F - Start a New File: Not Available In Built In Menu
This option causes the unit to start a new data file on the next scan without having to turn
scanning off. Using this option at the end of a site visit will cause all the data from that point on
to be in a new file so that on the next visit all the data in the file will be that collected between
the visits.