3300-0188-0M Rev L y apex Digital Scale Owner’s 26
#! (HOLD TIME)
This setting is used by the scale to hold a stable patient weight for a desired amount of time.
For example, if a value of 5 seconds is used, then when the scale locks onto a stable
patient weight, it will remain locked for 5 seconds before automatically releasing the weight.
The display will show 6/3 and the current setting. If the value displayed is acceptable,
press the
key. Otherwise, press the
×
or
Ø
keys to select a new value and then press
the
key to save it and proceed to the next prompt.
This setting controls the auto-locking feature. Allowable values for hold time are 0 to 20.
! (CALIBRATION)
The display will show ! and the current setting 06. If the scale has been previously
calibrated and you wish to skip calibration and proceed to the ;($ (Scale Model) prompt,
press the
key and the previous calibration will be retained.
To begin calibration, press the
×
key to select YES and then press the key. After
pressing the
key, the display will change to !#3.
!#3 (LOAD CALIBRATION WEIGHT)
With the display showing !#3, perform the following steps:
1. Make certain the scale platform is empty and free of debris.
2. Place the desired amount of calibrated test weights on the scale platform. A minimum
of
50% o
f scale's capacity is required. However, 70% to 100% is recommended.
3. Press the
key.
4. If the value displayed is acceptable, press the key again. Otherwise, determine the
exact amount of test weight placed on the scale platform and then using the
×
or
Ø
keys select the test weight amount.
5. Verify that the numbers selected are the same as the amount of the test weight and
then press the
key.
6. Starting at the left and preceding right, a series of dashes will appear on the display.
The dashes will stay on the display momentarily, then disappear, after which the displa
y
will proceed to the next prompt.
'0!3 (UNLOAD CALIBRATION WEIGHT)
After a moment, the display will change to '0!3.
1. Remove the test weights from the scale platform and then press the
key.
2. Starting at the left and preceding right, a series of dashes will appear on the display.
The dashes will stay on the display momentarily, then disappear, after which the
calibration factor will be saved and the display will proceed to the next prompt.
IMPORTANT: During the time the dashes are appearing on the display,
ensure t
hat the loaded (or empty) scale is stable.

Questions and Answers: