EasyManua.ls Logo

DeVilbiss DV63 Standard Plus - Patient Messages and Alerts; Notifications and Reminders; Device Status Messages; Service Codes and Troubleshooting Alerts

DeVilbiss DV63 Standard Plus
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EN - 48 SE-DV64-HE
PATIENT MESSAGES
Various messages may be displayed during your therapy; some include an audible alert
that is controlled by Master Volume Setting.
1. Notifications
These are friendly reminders to clean or replace components and, if enabled, are
generated as follows. Dismiss from Display by Button Press or Knob Movement.
Displayed Notication Recurrence interval
Air Intake Filter Due for Cleaning 50 hours of use (while breathing)
Mask, Tubing, and Chamber Due for Cleaning 35 hours of use (while breathing)
2. Messages
These identify conditions that may require action by you/your provider or are
informational only. Dismiss from Display by Button Press or Knob Movement.
Displayed Message Condition for Occurrence
Mask Off Please Check Mask Fit 10 seconds after mask off condition detected.
Auto-STOP Detected When Auto-STOP occurs.
Auto-START Detected When Auto-START occurs.
Wake-Up Alert <current time> When Wake-Up Alert activates (alarm clock).
Error with Code, Prescription Not
Updated
When SmartCode Rx is entered but incorrect.
Card Detected Please Wait … When SD card is detected. Do not remove SD
card while this message is displayed.
Card Removed Send to Provider When SD card is removed after data transfer.
Card Error Contact Provider If SD card error condition detected.
Settings Not Updated Contact
Provider
When prescription change via SD Card fails.
Card Transfer Successful When valid SD Card inserted and data transfer
complete.
Card Detected Ready for use When valid SD Card is inserted.
Firmware Update Please Wait … When SD Card containing rmware update is
inserted.
3. Add-On Device Messages
These messages may be displayed when Add-On items are in use, Dismiss from Display
by Button Press or Knob Movement.
Displayed Message Condition for Occurrence
Oximeter Logging Started
SpO2 _____
Pulse Rate _____
When Oximeter rst detected and logging
starts
Oximeter Logging Stopped When Oximeter signal is lost and logging
stops
Oximeter Finger Probe Off When Oximeter nger probe off condition
detected
Oximeter Finger Probe On
SpO2 _____
Pulse Rate _____
When Oximeter nger probe off condition
corrected
Wireless Modem Low Battery, Please
Recharge
When DV6WM is paired and battery level is
low
4. Service Codes
These are divided into Critical or Non-Critical. Critical service codes put the unit in a fail-
safe (blower off) state.
Displayed Message Action
Critical Service Code E0X - Where (X) is a
number or letter
NOTE – Code E01 is always shown in
English since language setting is unknown
Contact your equipment provider to
return unit for service. Message remains
on Display and cannot be dismissed.
Service Code E8X (non-critical - Where (X)
is a number or letter
Contact your equipment provider.
Dismiss from Display by Button Press or
Knob Movement.
NOTE- Service Code alerts are unaffected by notication enable/disable setting.