EasyManua.ls Logo

DeVilbiss DV63 Standard Plus - Supplemental Oxygen Use; Oxygen Use Warnings; Oxygen Connection Methods

DeVilbiss DV63 Standard Plus
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
EN - 51SE-DV64-HE
1. Place optional oxygen
adapter into air supply port.
3. Attach oxygen
tubing to the adapter.
2. Insert CPAP tubing
into adapter.
1. Attach oxygen tubing directly to the oxygen port on
the mask.
OR
Always turn on CPAP before
turning on oxygen ow.
Always turn off oxygen ow
before turning off CPAP.
SUPPLEMENTAL OXYGEN
WARNING-OXYGEN USE
Oxygen supports combustion. To avoid possible physical injury, do not smoke while using this device with supplemental oxygen. Do not use this device near hot objects,
volatile substances or sources of open ames.
Always turn on the device before turning on the oxygen source. Turn off the oxygen source before turning off the device. Never allow the oxygen source to run
continuously while connected to the device if the device is not in use. If the device is not being used, turn off the oxygen ow.
CPAP pressure must be set higher than 8 cmH2O for supplemental oxygen use to prevent patient breathing pattern from forcing oxygen back into the CPAP.
At a xed ow rate of supplemental oxygen, the inhaled oxygen concentration will vary depending on the pressure setting, patient breathing pattern, mask selection, and leak rate.
This warning applies to most types of CPAP devices.
CAUTION
Oxygen is a prescribed gas and should only be administered under the supervision of a physician.
The ow setting of the oxygen source must be specied by a physician.
Maximum oxygen pressure is 50 psi. Maximum oxygen ow rate is 10 lpm.
If your physician prescribed supplemental oxygen it can be added one of two ways:

Other manuals for DeVilbiss DV63 Standard Plus

Related product manuals