EN - 51SE-DV64-HE
1. Place optional oxygen
adapter into air supply port.
3. Attach oxygen
tubing to the adapter.
2. Insert CPAP tubing
into adapter.
1. Attach oxygen tubing directly to the oxygen port on
the mask.
OR
Always turn on CPAP before
turning on oxygen ow.
Always turn off oxygen ow
before turning off CPAP.
SUPPLEMENTAL OXYGEN
WARNING-OXYGEN USE
• Oxygen supports combustion. To avoid possible physical injury, do not smoke while using this device with supplemental oxygen. Do not use this device near hot objects,
volatile substances or sources of open ames.
• Always turn on the device before turning on the oxygen source. Turn off the oxygen source before turning off the device. Never allow the oxygen source to run
continuously while connected to the device if the device is not in use. If the device is not being used, turn off the oxygen ow.
• CPAP pressure must be set higher than 8 cmH2O for supplemental oxygen use to prevent patient breathing pattern from forcing oxygen back into the CPAP.
• At a xed ow rate of supplemental oxygen, the inhaled oxygen concentration will vary depending on the pressure setting, patient breathing pattern, mask selection, and leak rate.
This warning applies to most types of CPAP devices.
CAUTION
• Oxygen is a prescribed gas and should only be administered under the supervision of a physician.
• The ow setting of the oxygen source must be specied by a physician.
• Maximum oxygen pressure is 50 psi. Maximum oxygen ow rate is 10 lpm.
If your physician prescribed supplemental oxygen it can be added one of two ways: