Filter Specifications
Standard Filter ........................................................................................................................................................................................................................................... > 3.0 micron particles
Optional Fine Particle Filter ....................................................................................................................................................................................................................... > 0.3 micron particles
Dynamic Short-Term Pressure Accuracy with and without DV6HH Humidifier (Per ISO 17510-1:2007)
Waveform Volume [mL] Breath Rate [min-1]
Set Pressure
[cmH
2
O]
Pressure Accuracy
CPAP only
[Pk-Pk cmH
2
O]
Pressure Accuracy
CPAP with Humidier
[Pk-Pk cmH
2
O]
Sinusoidal cycle with
I:E = 1:1
500 10, 15, 20
3.0 0.5 0.5
7.0 0.5 0.5
12.0 1.0 1.0
16.0 1.0 1.0
20.0 1.0 1.0
Maximum Flow-Rate (per ISO 17510-1:2007)
Test Pressures
3.0 cmH
2
O 7.0 cmH
2
O 12.0 cmH
2
O 16.0 cmH
2
O 20.0 cmH
2
O
Measured pressure at the patient connection port (cmH
2
O) 2.9 6.6 11.4 15.0 19.0
Average ow at the patient connection port (l/min) 78.2 123.5 166.8 157.7 138.0
Long Term Static Pressure Accuracy ................................................................................................................................................................................................................... +/- 0.5 cmH2O
Additional Specifications
Equipment classification with respect to protection from electric shock ...........................................................................................................................................................................Class II
Degree of protection from electric shock ....................................................................................................................................................................................................Type BF Applied Part
Degree of protection against ingress of liquids .................................................................................................IP21 Ingress Protection – Protected against finger access to hazardous parts;
........................................................................................................................................................................................................................................................................................................
protected against vertically falling water drops
Mode of operation .......................................................................................................................................................................................................................................................Continuous
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous oxide. See instructions for use with supplemental oxygen.
US Patents
SmartCode Technology is protected under .................................................................................................................................................................................................. US Patent 8649510
Approvals
Meets RTCA/DO-160 .................................................................................................................................................... DO-160 Section 21 Category M for battery operation only Airline Use