16 Technical Information | Stelo Glucose Biosensor User Guide 105
Immunity test Transmitter compliance level
Magnetic Field (50Hz)
IEC 61000-4-8
30 A/m
Electrical Fast
Transient/Burst
IEC 61000-4-4
N/A
Surge
IEC 61000-4-5
N/A
Voltage Dips and
Interruptions
IEC 61000-4-11
IEC 60601-1-11
N/A
Conducted Fields
Disturbance
IEC 61000-4-6
N/A
Radiated Fields Disturbance
IEC 61000-4-3
10 V/m
at 80 MHz to 2700 MHz (AM Modulation)
Radiated and Conducted
Fields
Aircraft use
Meets RTCA /DO-160 edition G Section 20,
Category T