ROUTINE UTILIZATION & MEASUREMENT
37
5.2.7. Using Pre-diluted Mode
Pre-diluted measurement mode allows to measure insufficient sample for normal
mode, or if some parameter is out of the linearity range (WBC = 300 x 10
3
cells/µl).
Perform an external pre-dilution of the sample using clean isotonic saline solution, or
diluent reagent. Dilute the sample to 1:5 ratio (1 part sample to 5 part diluent), using
a clean sample vial. Mix it well.
To perform the analysis on pre-diluted sample:
1. In Main menu select Measure
2. New
3. In Options, select Pre-diluted mode
4. Put pre-diluted sample into the sample adapter
5. Press START button. The analyzer will automatically calculate the results with
the 1:5 pre-dilution factor.