7.2.2. Calibration by measurement
The analyzer can run calibrator, and perform calculation of factors automatically. User can
decide the number of measurements to use for calibration.
Maintenance
Calibration / Measurement
Prior to starting the calibration
measurements, you have to define some
basic parameters for the upcoming
measurements.
As the next step, target values from the assay sheet of the calibrator or control must be
entered before measurement.
Maintenance
Calibration / Measurement
Prior to starting the calibration
measurements, you have to define some
basic parameters for the upcoming
measurements.
Target values for calibrated parameters can be set within the following ranges:
Parameter
Low limit
High limit
WBC
1.0
30.0
RBC
1.00
8.00
HGB g/l
30
300
MCV
50
120
RDW CV
10
50
PLT
30
800
MPV
5
15
PDW CV
5
50
HCT
0.1
0.6
PCT
0
2
Table 7. Calibration target ranges
When all parameters are set, press Accept key.