7.4. Internal Count And A/D Count Display :
If the SPEC 25, bit 2 is set, then the [SPAN] Switch must be "ON" to enter this mode.
1
Enter [0] [0] [9] while depressing the [REZERO] Key. Unit Weight Window will display
the Internal Count and the Quantity Window will display the A/D Count.
2
Press [MODE] Key to exit from Program Mode to Weighing Mode.
3
Depressing the [SPAN] Switch returns scale to the Weighing Mode.
7.2. Spec 141 Setting
Spec 141 Setting:
Spec 141(Customer Specifications) can be accessed from the weighing mode.
1. Enter [1][4][1] while depressing the [REZERO] key.
2. [++
++
] key increases to the next SPEC number and also stores temporarily the SPEC data in the RAM location.
3. Enter 1011 as the new value for SPEC01 using the [Numeric] keypad
4. [C] key clears the [Numeric] entry.
5. [++
++
] key increases to the next SPEC number.
6. [−−
−−
] key decreases to the previous SPEC number.
7. [−−
−−
]key decreases from SPEC 01 to SPEC 00.
8. [−−
−−
] key decreases the SPEC number from SPEC00 to SPEC18
9. [❋] stores the new SPEC values to the NOV-RAM and exits from the SPEC setting mode.
10. Press [MODE] key to escape from maintenance mode to weighing mode.
7.3. Spec 142 Setting
Spec 142 Setting: To access the Spec 142 (W & M Spec) mode the [SPAN] Switch must be "ON".
The rest of the procedure is the same as Spec 141 setting.
See page 10 for location of the span switch
1 Depress the [SPAN] Switch. The S-On message comes on.
2 Enter [1][4][2] while depressing the [REZERO] key.
3 [++
++
] key increases to the next SPEC number and also stores temporarily the SPEC data in the RAM location.
4 Enter, (example, 1011) as the new value for SPEC21 using the [Numeric] keypad
5 [C] key clears the [Numeric] entry.
6 [++
++
] key increases to the next SPEC number.
7 [−−
−−
] key decreases to the previous SPEC number.
8 [−−
−−
]key decreases from SPEC 21 to SPEC 20.
9 [−−
−−
] key decreases the SPEC number from SPEC20 to SPEC36
10 [❋ PROG] key stores the new SPEC values to the NOV-RAM and exits from the SPEC setting mode.
11 Press [MODE] Key to exit from Spec Setting mode. The display shows S-On indicating that the [SPAN]
Switch is ON.
12 Depressing the [SPAN] Switch returns scale to the Weighing Mode.