EasyManua.ls Logo

DLab D3024 - Routine Inspection; 10 Troubleshooting; Possible Problems and Solutions

DLab D3024
27 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
USER MANUAL
18
Seal ring 1 2 3
Seal ring 2 3
Figure 9-1 Seal rings of rotor
9.3.2 Replacement methods
1Clean the seal ring slot with neutral detergent solution and make it dry.
2Evenly coated with glue (501) in the seal ring slot and keep the seal ring into slot, press evenly to make
it contact enough with the slot bottom and bond firmly.
3Place for 20 minutes and waiting for the glue to completely solidified.
9.4 Routine inspection
(1) Check that if the centrifuge is on a firmflat and level table, ensure the four feet stand on the table
firmly.
(2) Check if the centrifuge grounded properly: Use multi-meter to check if is short circuit between the
power cord grounding pin and the motor shaft. If yes, indicating grounded properly; if is open circuit, need
to check failure reason first and make troubleshooting before use.
10. Troubleshooting
10.1 Possible problems and solutions
This centrifuge has a self-diagnostic function. If a problem occurs, an error/warning code will be displayed
on the time display screen and the operator can determine the malfunction with the warning code below.
Symptom
Causes
Solutions
Nothing appears on the screen when the
POWER is turned on.
·
Building power circuit
breaker trips.
·
the fuse was blown out.
·
Remove the trouble and
turn on the POWER.
·
Replace the fuse.

Related product manuals