refrigeration is not running.
8.1.4 Start the operation
(1)Press button to start running
⚫ The door should be locked before rotor starts rotating.
⚫ Timer will operate once the speed setting value is reached, the screen displays the remaining run time.
(2) View and modify the operation programs
⚫ Pressing button , returns the display to the program interface and displays settings programs. Press
the select button to the desired program. When flashing, rotate parameter button to modify
values. Release the button after 5 seconds, and the centrifuge will return to normal operation mode and
run according to the new value.
⚫ If the set time value has been modified, the operation time is not affected and will continue.
(3)Warning display
⚫ If an error occurs during the operation, the centrifuge will brake to stop automatically, and display the
error code on the time/display area. The error code can be checked in the table 10-1, and corrective
actions can be applied accordingly.
8.1.5 End the operation
(1)The centrifuge will brake when it reaches the setting time or button is pressed.
⚫ When the rotor stops rotating, centrifuge will start beeping to alert the operation has finished.
(2)Open the door
⚫ The door can be released automatically when the operation has stopped (D3024).
⚫ The centrifuge keeps the door close when operation has stopped ( D3024R ).
⚫ With the door closed, you are able to press the button to open it.
⚫ After ending the operation, the program will store the setting parameters of this operation, and will
recall these parameters when restart the program.
(3)Open the door and take out the rotor and samples.
8.2 RCF Operation
(1)Turn on the power switch.
(2)Set a RCF (Relative Centrifugal Force) value.