11.2.4 Condition and life expectancy of tubes
The life expectancy of plastic tubes depends on the characteristics of samples, speed of the rotor used, and
temperature applied, and so on. When the plastic tubes are used for centrifuge of ordinary aqueous samples
(pH between 5 and 9), their life expectancies are defined as follows.
Be operated at the maximum speed:
High quality tubes (PA、PC、PP): 30-50 operations
Ordinary tubes(PA、PC、PP): around 10 operations(Using in low speed can extend the tube life).
Life expectancy of tubes also depends on the pretreatment conditions such as cleaning and sterilization,
lifetime can be cut down.
Notice: Do not use damaged or cracked tubes.
12. Calculate Relative Centrifuge Force(RCF)
Relative Centrifuge Force (RCF) can be determined with the following calculation formula.
RCF=1.118×r×n
2
×10
-5
R—rotating radius, unit: cm; n—rotating speed, unit: rpm
13. Ordering information
High Speed Micro-Centrifuge,with AS24-2 rotor kits, US
plug, 110V/50Hz/60Hz
High Speed Micro-Centrifuge,with AS24-2 rotor kits, Cn plug,
220V/50Hz/60Hz
High Speed Micro-Centrifuge,with AS24-2 rotor kits, Euro
plug , 220V/50Hz/60Hz
High Speed Micro-Centrifuge,with AS24-2 rotor kits, UK
plug, 220V/50Hz/60Hz
High Speed Refrigerated Micro-Centrifuge,with AS24-2 rotor
kits, US plug, 110V/50Hz/60Hz
High Speed Refrigerated Micro-Centrifuge,with AS24-2 rotor
kits, Cn plug, 220V/50Hz/60Hz
High Speed Refrigerated Micro-Centrifuge,with AS24-2 rotor
kits, Euro plug, 220V/50Hz/60Hz
High Speed Refrigerated Micro-Centrifuge,with AS24-2 rotor
kits, UK plug, 220V/50Hz/60Hz
Rotor kits with cover, fixing clips and O'ring, 15000rpm,
2ml*24, used with D3024 & D3024R