USER MANUAL
20
10.3 Replacement of fuses
1There are two fuses of D3024, 250V, 5A time-delay type, size: Ф5×20. There are two fuses of
D3024R, 250V, 10A time-delay type, size: Ф5×20.
2The fuse holder is located in the power inlet. Pull out the fuse holder from power inlet and replace
the fuses if necessary.
11. Instructions of rotor and tube
CAUTION
Read the instructions thoroughly, correct use rotor.
Do not exceed the allowable maximum speed of rotortube and adapters etc., be care that the allowable
maximum speed of some adapters are lower than the rotor
s maximum speed.
11.1 The rotor instructions
11.1.1 Rotor structure
Knob Locking nut Cover
Seal ring Shaft hole
Figure 11-1 The rotor structure
11.1.2 Available rotors and adapters
All rotors are used for bio-safe when the rotor lid was tightened with the rotor, centrifuge tubes will be