Main interface of D3024R model is as figure 6-3. The speed is set to be 15000rpm, temperature of
centrifugal chamber is 4℃, and the running time is 12 minutes. When speed symbol is rotating, indicating
the centrifuge is running, the rotation is faster, the speed is higher. When temperature symbol is blinking,
indicating refrigeration is running, no blinking means refrigeration stop working.
Time symbol displays the ratio of working to time setting. The total time setting is divided into 10
scales.
7. Rotor Preparation
7.1 Prepare the samples
7.2 Inject the samples into tubes.
CAUTION:
⚫ Do not overload samples into the centrifuge which will cause leaking.
⚫ Do not exceed the actual capacity allowed in the user manual.
7.3 Keep the tubes balance
⚫ Although the centrifuge can accept sample balancing by eye, we recommend that you keep this
centrifuge in a well-balanced condition to extend its life expectancy.
⚫ Never intentionally run the centrifuge under unbalanced condition even though the allowable
imbalance is not exceeded.
7.4 Inspect the rotor
Check the rotor for corrosion or scratches before using.
CAUTION:
⚫ If any abnormality such as corrosions or scratches are found, stop using the rotor and contact our
service center.
⚫ Only manufacturer
‟
s rotors must be used with the unit.
7.5 Symmetrically load centrifuge tubes in rotor
CAUTION:
⚫ Make sure the rotor lid is securely fixed on the rotor, as well as the rotor and shaft are tightened.