The lever is hidden under the right bottom of the centrifuge.
Pull the lever to the right and release the lock. Then open the lid.
11. Introduction to rotor and centrifuge tube
Caution
●Carefully read the user manual and correctly install and use the rotor correctly.
●Don‘t exceed the maximum allowed speed of the rotor, test tube and adaptor. The
maximum allowed speed by certain adaptors is lower than the maximum speed of the
rotor and check before operation.
11.1.Rotor description
11.1.1. Rotor structure
Figure 11.1 Rotor structure
11.1.2. Rotor and adaptor
The rotors and adaptors which are suitable for use with this centrifuge are listed as follows:
Rotor
Model
Rotor
No.
Centrifuge tube
Adaptor
Max
speed
(rpm)
Max RCF
(×g)
Balance Tolerance(*)
Mass
Volume(**)
A6-15P
01
1.5/2.0ml micro
centrifuge tube
A2P17
5000
1817
2.0g/tube
φ13x75 Blood
collection tube
A5P17
5000
2236
2.0g/
tube
15ml tube
5000
2600
2.0g/
tube
φ13x100 mm tube
5000
2600
2.0g/
Spring Locking Nut
Rotor Body Drive Shaft Tube Cavity