3.2. Transport and storage conditions
(1) Range of ambient temperature: -40℃-55℃.
(2) Range of relative humidity: ≤93%.RH
(3) The centrifuge must remain upright while in transit, suitably protected using wooden Kart box
(4) Lift the centrifuge by the chassis only.
(5) Pay attention to the centrifuge’s weight while in transit (see “1.Performance indicators”).
.
4. Installation
Users must strictly comply with the installation instructions contained in this chapter.
Be advised! Remove the rotator before moving the centrifuge.
Warning
Improper power connection might damage the centrifuge.
Before connecting the power supply, please check the power supply for
compliance with the requirements.
4.1. Mounting position
(1) This centrifuge must be mounted on a solid, flat and tabletop with contact between the four feet
of the centrifuge and the tabletop. Don’t mount the centrifuge on any sliding tabletop,
otherwise significant vibration might occur. Carefully place the centrifuge to avoid damage.
(2) The ideal ambient temperature is 20°C ±5°C and the ambient temperature should not be more
than 30℃. Avoid direct sunlight on this centrifuge.
(3) Place the centrifuge properly and keep >30cm space from the rear and >10cm space on both
sides to ensure effective air cooling.
(4) Ensure that there is no water leakage/ heat loss near the centrifuge as it may cause the rise in
temperature and thereby leading to centrifuge failure.
4.2. Connection between power cable and ground wire
Warning
Don’t touch the power cable with wet hands and avoid electric shocks.
Ensure that the centrifuge is well grounded.