1. Performance indicators
Maximum rotation speed
5,000rpm (300-5,000rpm)
step: 10rpm
Maximum relative centrifugal
acceleration
2,600×g, step: 10×g
Capacity
15ml×6
Timing
30s-99 min; HOLD (continuous running)
Drive motor
Brushless motor
Safety performance
Door lock, overspeed, and onboard diagnosis system.
Power
Single phase, 100-240V, 50/60Hz, 2A
Dimensions (mm)
(Width) 300× (depth) 240× (height) 180
Weight
5.2kg
Acceleration and deceleration time
35s↑40s
Noise
≤56dB
Other functions
Rotation speed/RCF switch, sound notification, 2-level
deceleration; predefined and customizable programs.
2. Conformity to standards
The centrifuge’s structure conforms to the following safety standards:
EN 61010-1
EN 61010-2-10
UL 3101-1
The centrifuge’s structure conforms to the following electromagnetic compatibility standard:
EN 61326-1
Compliant with the following EU standards:
EMC standard: 89/336/EWG
Mechanical Design: 73/023/EWG
3. Environmental requirements
3.1. Basic operating requirements
(1) Power supply: single phase, 100-240V, 50/60Hz, 2A, sinusoidal waveform.
(2) Ambient temperature: 2~40℃.
(3) Relative humidity: ≤80%.RH
(4) No vibration or air flow present nearby that might affect performance.
(5) No conductive dust, explosive gas or corrosive gas exists in ambient air.