notify the service representative about the details.
⚫ Never touch the power cable/switch with wet hands and avoid electric shocks.
⚫ As a safety measure, ensure to maintain at least 30cm distance from the centrifuge while in
operation.
⚫ For safety reasons, when the centrifuge is running, the personnel should maintain a 30cm distance
from the centrifuge.
⚫ Never open the lid while the rotor is in operation.
⚫ The centrifuge should be opened for repair/ dismantling by trained staff only.
Caution
⚫ Ensure that the centrifuge is stably placed on level and solid work bench before any operation.
⚫ Ensure that the angle between the outer lid and housing is larger than 70° while opening the lid.
⚫ Never place hands/any other things in between the outer lid and the housing.
⚫ Don’t move or lean against the centrifuge if it is in operation mode.
⚫ If there is any liquid found in the centrifuge, ensure to wipe it OFF with a cloth in time to prevent
sample contamination.
⚫ Ensure that the centrifuge chamber is left clean and free of any foreign objects/ tube fragments
before each and every operation
⚫ Reminders about rotor:
⚫ Check and ensure that the rotor surface is free of corrosions/ damages before its operation.
(1) The set rotation speed of the centrifuge should not exceed the allowed minimum speed of
the rotor assembly and accessories (rotor and adapter) and ensure to run the centrifuge under
minimum allowed speed.
(2) Don’t exceed the allowed amount of imbalance.
(3) The centrifuge tubes used should be within their allowed capacity.
(4) If the rotor has a lid, ensure to tighten it before operation.
(5) Use genuine accessories only.
⚫ If there is any abnormality or strange noise observed in operation, please stop the centrifuge
operation and contact the service center and intimate the failure code immediately.
⚫ Earthquake might cause damage to the centrifuge. If any abnormality occurs, please contact the
service center.