- 26 -
Symptom
Possible error
codes
Check/Action Remedy
Alarm, too low pres-
sure due to a reject
valve that does not
close.
13
Alarm, too low
pressure due to an
inlet valve that does
not open - either the
valve is faulty or not
energized.
14
Start the water purier. •
Check if it produces any
pure water or if the ow is
very low.
Measure the water ow in •
the reject hose.
When the motor is running, •
check that the inlet valve is
energized.
Check if any water ows •
in the drain hose when the
product is ushing. If water
ows, the inlet valve is
open.
Verify that the water purier •
is disconnected and then,
with care, slightly open one
of the lter containers. Re-
connect the product. Check
if any water comes out of
the lter container. If so, the
inlet valve opens properly.
water spillage, be
ready to unplug the
product.
If the reject valve remains in a ushing
position - change the reject valve.
In this case the reject ow will be > 5 l/minute.
If the inlet valve is not energized (no power to
the inlet valve), a wiring fault or a soldering
of a contact on the PCB might be the cause.
Replace the inlet valve if no water ows from
the lter container or in the drain hose (the
inlet valve does not open).
TROUBLESHOOTING