Ceiling and wall attachment
Dr. Mach
Lamps and Engineering
Static inspection
Note:
The static (structural) inspection must be carried out before the installation of the
ceiling or wall anchorage!
- The strength of the construction must be designed, checked and certified by a structural
engineer.
- The respective regional construction regulations that apply must be followed.
- If a wrong hole is drilled by mistake, e.g. drilling of a reinforcement rod, the structural engi-
neer who is responsible must be contacted, since adequate static load distribution in the cei-
ling may have been endangered.
Declaration of acceptance:
It is hereby certified that the supporting ceiling / wall and the ceiling anchoring / wall anchoring
is safe and adequately strong.
Project: ____________________________________________________________________
____________________________________________________________________________
____________________________________________________________________________
Anchoring (please check the one that is applicable)
- with dowels authorized by construction authority
- with counter-plate
- other
Location: ____________________________________________
Signature / Stamp: (structural engineer / construction authority)
59500001 Edition 06 11.03.2003 / Bak Page 5/48