46 Instructions for use | LAR 8000
en | Preparations for use
4. Plug the adapter of the dosage tester into the plug-in connector of the lung
demand valve.
The rebreather is prepared for testing the dosage.
4.2.2.2 Testing with dosage tester (before the dive)
Prerequisites
The rebreather is ready for the dosage test
Work equipment
Rebreather
Dosage tester
WARNING
Danger to life
If a dosage tester is used that is intended for a different nitrox mixture, the dosage
cannot be tested correctly.
Only use the dosage tester intended for the nitrox mixture in use.
1. Slowly open the valve of the mixed-gas cylinder.
2. Set the rotary switch to mixed operation (rotary switch is vertical and in the
housing).
3. When using a dosage tester with colored marking, the dosage must be within
the marked range.
Permissible dosage for nitrox B mixture: 5.0 to 6.5 L/min
Permissible dosage for nitrox C mixture: 10.0 to 13.0 L/min
Permissible dosage for nitrox D mixture: 16.0 to 19.0 L/min
If the dosage is outside the tolerance: Do not use rebreather and submit for
repair. Observe regulations and instructions for diving with oxygen and mixed
gas!
4. Close the mixed-gas cylinder valve.
45273
1 Lung demand valve connection 3 Dosage tester case
2 Dosage tester adapter
1
3
2