REPLACEMENT PROCEDURES (continued) NM2C
4.21.9 Reinstall the Datagrip and remote display assembly to the display
arm and tighten its mounting screw to a torque of 4 foot pounds.
Replace the plastic cap covering the mounting screw recess.
4.21.10 Route the cables as shown in the applicable illustration. For the
adjustable arm, coil the excess cable in the patient IV looms. For the
short arm, coil the excess cable in the ¾ in. cable clamp on the
underside of the ventilator box.
4.21.11 Connect the remote display and Datagrip cables to their ports on the
underside of the ventilator box. Verify that the arm and the Datagrip
have full range of motion with no binding caused by the cables.
Position the cables in their clamps as needed.
4.21.12 If tension adjustment is needed on the adjustable display arm hinge,
refer to the adjustment procedure in Section 5 of this manual.
4.21.13 Restore power to the machine and verify that the Datagrip and
remote display are working properly.
4.21.14 Perform the PMS Procedure given in Section 6.
SUPPORT PLATE
PATIENT LINE
BOOM ARM
1/4 - 20 X 7/8 IN. BUTTON
HEAD SOCKET SCREW
AND LOCKWASHER (4X)
3/8 - 24 X 2 3/4 IN.
FLAT HEAD
SOCKET SCREW (2X)
DISPLAY
ARM
1/4 - 20 X 1 3/4 IN.
SOCKET HEAD
CAP SCREW (2X)
SV20503
Figure 4-21C: OPTIONAL PATIENT LINE BOOM ARM ASSEMBLY DETAILS
Rev. B
4-66