EasyManua.ls Logo

Dräger Narkomed 2C - Bellows Drive Gas Leak: Adult

Dräger Narkomed 2C
274 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PMS PROCEDURE (continued) NM2C
6.21.19 Using a stopwatch, time the inspiratory phase.
( ) 6.21.20 What is the inspiratory time? ___ seconds (1.8-2.2)
6.21.21 Using a stopwatch, time the expiratory phase.
( ) 6.21.22 What is the expiratory time? ___ seconds (3.6-4.4)
** 6.21.23 Press and hold the EXTENDED RANGE switch and scroll the I:E ratio
dial counter-clockwise and verify the extended I:E values increment
(2:1, 3:1 and 4:1); return the I:E ratio to 2:1.
** 6.21.24 Using a stopwatch, time the inspiratory phase.
( ) ** 6.21.25 What is the inspiratory time? ___ seconds (3.6 - 4.4)
** 6.21.26 Using a stopwatch, time the expiratory phase.
( ) ** 6.21.27 What is the expiratory time? ___ seconds (1.8 - 2.2)
( ) 6.21.28 Adjust the FREQUENCY and I:E RATIO through the following settings
and verify that the ventilator cycles properly:
FREQ. I:E RATIO FREQ. I:E RATIO FREQ. I:E RATIO
11 1:1 44 1:2.5 77 1:4
22 1:1.5 55 1:3 88 1:4.5
33 1:2 66 1:3.5 99 1:4.5
6.22 BELLOWS DRIVE GAS LEAK: ADULT
6.22.1 Remove the ventilator hose from the VENTILATOR HOSE terminal.
6.22.2 Attach a test terminal to the bellows assembly ventilator hose terminal.
6.22.3 Connect a flowmeter test stand (P/N S000081) to the test terminal.
6.22.4 Set the FREQUENCY to 1 BPM.
6.22.5 Set the I:E RATIO to 1:1.
6.22.6 Set the INSPIRATORY FLOW to the maximum.
( ) 6.22.7 What is the flow indicated during the inspiratory phase? ___ (<50 ml)
6.22.8 Remove the test terminal and flowmeter test stand.
6.22.9 Turn the ventilator and all ventilation alarms OFF.
Rev. Q
6-46
RETURN TO CD-ROM TABLE OF CONTENTS
RETURN TO THIS MANUAL'S TABLE OF CONTENTS

Table of Contents

Related product manuals